DISSOLUTION : 

DISSOLUTION SWAPNA MASURKAR ARDL - I

WHAT IS DISSOLUTION? : 

WHAT IS DISSOLUTION? the process by which a solid or liquid forms a homogeneous mixture with a solvent Tablet Dissolution is a standardised method for measuring the rate of drug release from a dosage form

FUNCTIONS OF DISSOLUTION : 

FUNCTIONS OF DISSOLUTION • Optimisation of therapeutic effectiveness during product development and stability assessment • Routine assessment of production quality to ensure uniformity between production lots • Assessment of ‘bioequivalence’, that is to say, production of the same biological availability from discrete batches of products from one or different manufacturers • Prediction of ‘in-vivo’ availability, i.e. bioavailability (where applicable)

TYPES OF APPARATUS : 

TYPES OF APPARATUS BASKET APPARATUS PADDLE APPARATUS RECIPROCATING CYLINDER FLOW-THROUGH CELL PADDLE OVER DISK CYLINDER RECIPROCATING HOLDER

BASKET APPARATUS : 

BASKET APPARATUS VESSEL A MOTOR, METALLIC DRIVE SHAFT CYLINDRICAL BASKET

VESSEL : 

VESSEL CYLINDRICAL WITH HEMISPHERICAL BOTTOM CAPACITY : 1 TO 4 LITERS FOR 1 LIT : HEIGHT – 160 TO 210 MM INNER DIAMETER – 98 TO 106 MM COVER MAY BE USED TO RETARD EVAPORATION

SHAFT : 

SHAFT POSITIONED IN SUCH A WAY THAT ITS AXIS IS NOT MORE THAN 2MM FROM VERTICAL AXIS OF THE VESSEL SHOULD ROTATE SMOOTHLY WITHOUT SIGNIFICANT WOBBLE

BASKET : 

BASKET

PADDLE APPARATUS : 

PADDLE APPARATUS METALLIC, SUITABLY INERT, RIGID BLADE AND SHAFT COMPRISING OF SINGLE ENTITY SINKERS (A SMALL, LOOSE PIECE OF NON-REACTIVE MATERIAL) MAY BE ATTACHED TO DOSAGE UNIT TO AVOID FLOATING

MEDIA FOR DISSOLUTION : 

MEDIA FOR DISSOLUTION POINTS TO BE REMEMBERED WHILE SELECTING A MEDIUM VOLUME DEAERATION EXAMPLES OF TYPICAL MEDIA

POINTS TO BE REMEMBERED WHILE SELECTING A MEDIUM : 

POINTS TO BE REMEMBERED WHILE SELECTING A MEDIUM PHYSICAL AND CHEMICAL DATA SOLUBILITY AND STABILITY AS A FUNCTION OF pH FOR POORLY SOLLUBLE – SURFACTANTS SUCH AS SLS, POLYSORBATE RELEASE MECHANISM

VOLUME OF MEDIUM : 

VOLUME OF MEDIUM 500 TO 1000 ml 900 ml IS GENERALLY USED CAN BE RAISED TO 2 TO 4 LIT

DEAERATION : 

DEAERATION AIR BUBBLES CAN : INTERFERE WITH THE RESULT CAN CAUSE PARTICLES TO CLING TO THE APPARATUS AND VESSEL WALLS DEAERATION CAN BE DONE BY: HEATING FILTERING VACUUM

EXAMPLES OF TYPICAL MEDIA : 

EXAMPLES OF TYPICAL MEDIA WATER DILUTE HCl BUFFERS OF pH RANGE 1.2 TO 7.5

Slide 16: 

APPARATUS AGITATION TIME POINTS SAMPLING FILTERS ASSAY

APPARATUS : 

APPARATUS FOR SOLID ORAL DOSAGE FORMS, BASKET OR PADDLE METHOD IS USED IN CASE OF BASKET, MESH SIZE RANGING FROM 10 TO 80 CAN BE USED SINKERS CAN BE ADDED IN PADDLE METHOD

AGITATION : 

AGITATION RANGE : 25 TO 100 RPM FOR BASKET METHOD : 100 RPM FOR PADDLE METHOD : 50 TO 75 RPM BELOW 25 RPM : INCONSISTENCY OF HYDRODYNAMICS ABOVE 150RPM :TURBULANCE

TIME POINTS : 

TIME POINTS FOR IMMEDIATE RELEASE : 15 TO 60 MINS FOR EXTENDED RELEASE : AT LEAST THREE TEST TIME POINTS ARE SELECTED

SAMPLING : 

SAMPLING MANUAL : USING PLASTIC OR GLASS SYRINGE, A STAINLESS STEEL CANNULA AUTOSAMPLING : BEST FOR SEVERAL TIME POINTS CAN BE SEMI OR FULLY AUTOMATIC

FILTERS : 

FILTERS USED TO PREVENT UNDISSOLVED DRUG PARTICLES FROM ENTERING THE ANALYTICAL SAMPLE USED TO REMOVE INSOLUBLE EXCIPIENTS WHICH MAY CAUSE TURBIDITY PORE SIZE MAY RANGE FROM 0.45 TO 70 µM

APPARATUSAGITATIONTIME POINTSSAMPLINGASSAY : 

SPECTROPHOTOMETRY OR HPLC SPECTROPHOTOMETRY IS PREFERRED BECAUSE : ANALYSIS IS SIMPLE AND FASTER FEWER SOLVENTS ARE USED APPARATUSAGITATIONTIME POINTSSAMPLINGASSAY

SPECTROPHOTOMETRY : 

SPECTROPHOTOMETRY CELLS WITH PATHLENGTHS 0.02 TO 1 CM CAN BE USED ALIGNMENT OF CELLS MUST BE PROPER SMALLER PATHLENGTHS ARE USED TO AVOID DILUTING THE SAMPLE A TYPICAL BLANK, STANDARD AND SAMPLE ARE ANALYSED IN A SEQUENCE TYPICAL RSD IS NOT MORE THAN 2 %

HPLC : 

HPLC COMPATIBILITY OF DISSOLUTION MEDIA AND MOBILE PHASE MUST BE EXAMINED SINGLE INJECTION OF EACH VESSEL TIME POINT WITH STANDARDS THROUGH-OUT THE RUN CONSTITUTE A TYPICAL RUN