Two Models of Medical Error Reduction Programs in Radiation Oncology: 

Two Models of Medical Error Reduction Programs in Radiation Oncology by Ed Kline RadPhysics Services LLC Albuquerque, NM © RPS

Introduction: 

Introduction Patient safety Freedom from accidental injury due to medical care, or absence of medical errors1,2 or Absence of misuse of services3,4 In radiation oncology, variety of injuries and errors can occur in the diagnostic imaging or therapeutic treatment delivery processes 1 Hurtado M, Swift E, Corrigan JM, eds. Envisioning the National Health Care Quality Report. Washington, DC: National Academy of Sciences; 2001. 2 McNutt R, Abrams R, Arons D. Patient Safety Efforts Should Focus on Medical Errors. JAMA. 2002;287(15):1997-2001. 3 Department of Health and Human Services. The Challenge and Potential for Assuring Quality of Health Care for the 21st Century. Washington, DC: Department of Health and Human Services; 2000. 4 The President's Advisory Commission on Consumer Protection and Quality in the Health Care Industry. Quality First: Better Health Care for All Americans; 1998.

Introduction: 

Introduction This presentation describes the design, implementation, and results of two QA/medical error reduction programs Both programs are designed for Reducing preventable systems-related medical errors (i.e., sentinel events, “near misses”) Preventing violations of regulatory requirements (i.e., State/NRC) Ensuring compliance with recommended standards (i.e., JCAHO, ACR, ACRO, etc.)

History: 

History Institute of Medicine (IOM) report5 Focused a great deal of attention on the issue of medical errors and patient safety 44,000 to 98,000 deaths per year in U.S. hospitals each year as the result of medical errors 10,000 deaths per year in Canadian hospitals Exceeds annual death rates from road accidents, breast cancer, and AIDS combined in U.S. 5 To Err is Human: Building a Safer Health System. Institute of Medicine (IOM). The National Academies (11/29/99).

History: 

History IOM Costs6 Approximately $37.6 billion per year About $17 billion are associated with preventable errors Of that $17 billion, about $8 to $9 billion are for direct health care costs 6 To Err is Human: Building a Safer Health System. Institute of Medicine (IOM). National Academies (11/29/99).

History: 

History Federal initiatives7 taken by former President Clinton on 2/22/00 based on IOM recommendations8 Comprehensive strategy for health providers to reduce medical errors Creation of external reporting systems to identify and learn from errors so as to prevent future occurrences Creation of national patient safety center to set goals At least 50% reduction of errors over 5 years 7 Announced by President Clinton and senior administration officials in James S. Brady Press Briefing Room on February 2, 2000. 8 Recommendations issued in report entitled To Err is Human: Building a Safer Health System by the Institute of Medicine (IOM) of the National Academies (11/29/99).

History: 

History Key legislation Patient Safety Quality Improvement Act9 Certifies patient safety organizations in each State to collect data and report on medical errors State Patient Safety Centers In past 5 years, 6 states have enacted legislation supporting creation of state patient safety centers 5 of the 6 states now operate patient safety centers Separate mandatory reporting systems for serious adverse events Centers are housed within state regulatory agencies 9 Reducing Medical Errors, Issue Module, Kaiser EDU.org, Accessed through www.kaiseredu.org.

History: 

History Patient safety centers include10 The Florida Patient Safety Corporation The Maryland Patient Safety Center The Betsy Lehman Center for Patient Safety and Medical Error Reduction (Massachusetts) The New York Center for Patient Safety The Oregon Patient Safety Commission The Pennsylvania Patient Safety Authority 10 State Patient Safety Centers: A New Approach to Promote Patient Safety, The Flood Tide Forum, National Academy for State Health Policy, 10/04, Accessed through www.nashp.org.

History: 

History State reporting: mandatory vs voluntary11 Mandatory reporting: Colorado, Florida, Kansas, Nebraska, New York, Ohio, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Washington Voluntary reporting: District of Columbia, Georgia, New Mexico, North Carolina, Oregon, Wyoming Considering new legislation: Arizona, California, Maine Mandatory reporting but considering new legislation: Massachusetts, New Jersey 11 National Conference of State Legislatures, National Academy for State Health Policy, 12/03, Accessed through www.nashp.org.

History: 

History JCAHO revises standards Patient safety standards effective 7/1/01 Requires all JCAHO hospitals (5,000) to implement ongoing medical error reduction programs Almost 50 percent of JCAHO standards are directly related to safety12 12 Patient Safety - Essentials for Health Care, 2nd edition, Joint Commission on Accreditation of Healthcare Organizations. Oakbrooke Terrace, IL: Department of Publications, 2004.

History: 

History JCAHO’s sentinel event policy13 Implemented in 1996 Identify sentinel events Take action to prevent their recurrence Complete a thorough and credible root cause analysis Implement improvements to reduce risk Monitor the effectiveness of those improvements Root cause analysis must focus on process and system factors Improvements must include documentation of a risk-reduction strategy and internal corrective action plan Action plan must include measurements of the effectiveness of process and system improvements to reduce risk 13 Sentinel Event Policies and Procedures -  Revised: July 2002, Joint Commission on Accreditation of Healthcare Organizations, Accessed through www.jcaho.org/accredited+organizations/long+term+care/sentinel+events/index.htm.

History: 

History JCAHO’s Office of Quality Monitoring Receives, evaluates and tracks complaints and reports of concerns about health care organizations relating to quality of care issues Conducts unannounced on-site evaluations JCAHO and CMS agreement14 Effective 9/16/04 Working together to align Hospital Quality Measures (JC’s ORYX Core Measures and CMS’7th Scope of Work Quality of Core Measures) 14 Joint Commission, CMS to Make Common Performance Measures, Joint Commission on Accreditation of Healthcare Organizations, Accessed through www.jcaho.org/accredited+organizations/long+term+care/sentinel+events.

History: 

History CMS quality incentives15 Quality Improvement Organizations (QIOs) Contracted by CMS to operate in every State 67% of QIOs perform independent quality audits Premier Hospital Quality Initiative 3-year demonstration project recognizes and provides financial reward CMS partnership with Premier Inc., nationwide purchasing alliance Hospitals in top 20% of quality for specific diagnosis get financial reward Top decile gets 2% Diagnosis Related Group (DRG) bonus 2nd decile get 1% DRG bonus Hospitals performing below 9th and 10th decile baseline levels, DRG payments reduced 1% and 2%, respectively 15 Medicare Looks for Ways to Boost Quality Care Comments Sought on New Plan for Quality Improvement Organizations, Centers for Medicare & Medicare Services (CMS), Accessed through www.cms.hhs.gov.

History: 

History CMS quality incentives CMS consumer website Beginning in 4/05, hospital quality data available at www.HospitalCompare.hhs.gov or 1-800-MEDICARE Data indicators16 In 2006, hospitals reporting quality data to Medicare receive 3.7% increase in inpatient payments Non-reporters receive 3.3% increase Data covers 10 quality indicators for cardiology Plans are to expand into other disciplines 16 Medicare to Pay Hospitals for Reporting Quality Data, Modernhealthcare, accessed through www.modernhealthcare.com.

History: 

History CMS quality incentives Physician Voluntary Reporting Program17 Beginning in 1/06 Physicians voluntarily report information to CMS 36 evidence-based measures Information collected through Healthcare Common Procedure Coding System (HCPCS) CMS will provide feedback on physician’s level of performance 17 Medicare Takes Key Step Toward Voluntary Quality Reporting for Physicians, Centers for Medicare & Medicare Services (CMS), Accessed through www.cms.hhs.gov.

Radiation Oncology Errors: 

Radiation Oncology Errors Not well established No comprehensive numbers available for number of errors resulting in death18 Reported error rates range 0.1% to 0.2% of fields treated17 Studies not relying on self-reporting show actual rates of up to 3%17 18 French, J, Treatment Errors in Radiation Therapy. Radiation Therapist, Fall 2002, Vol. 11, No. 2; 2002.

Significant Medical Events in Radiation Oncology: 

Significant Medical Events in Radiation Oncology

Medical Error Rates in Radiation Oncology – Table 1: 

Medical Error Rates in Radiation Oncology – Table 1

Medical Error Rates in Radiation Oncology – Table 2: 

Medical Error Rates in Radiation Oncology – Table 2

Medical Error Rates in Radiation Oncology – Table 3: 

Medical Error Rates in Radiation Oncology – Table 3

Objective of a “QA/Medical Error Reduction Program”: 

Objective of a “QA/Medical Error Reduction Program” Objective was to provide a unified, total quality management and continuous improvement program for minimizing the occurrence of errors identified in the patient treatment process and regulatory arena. The program was designed for 17 geographically dispersed radiation oncology clinics located in 9 states of varying regulatory oversight and enforcement philosophy.

Design of QA/Medical Error Reduction Program: 

Design of QA/Medical Error Reduction Program Established a consistent set of of QA procedures for the 17 facilities following the strictest state requirements in which each facility resides. Analyzed the process of delivering radiation therapy to identify the steps used in all aspects of this modality. Developed a reporting codification system for errors detected, and the appropriate forms and procedures for reporting these errors. This includes a staging system for classifying the importance of an error.

Design of QA/Medical Error Reduction Program: 

Design of QA/Medical Error Reduction Program Provided an internal feed-back mechanism of corrective action to close the loop Independent review/recommendations for corrective action regarding all self-identified significant errors/violations Produced a quarterly report summarizing errors/violations Perform trend analysis of reported errors at center and company levels Recommended company wide corrective actions based on results of trend analysis

Slide26: 

Unintended Deviation Reporting Process

The Unintended Deviation System: 

The Unintended Deviation System The name was selected to convey an unintentional error discovered either by the one having committed the error or by another physician/staff member. Management emphasizes that self-identification and reporting of errors will not result in disciplinary action. Provides for identification, evaluation, and documentation of all errors within the process of radiation therapy delivery. Suggests possible causes and solutions for correction of individual errors as well as programmatic errors with discoverable trends.

Definition - Unintended Deviation : 

Definition - Unintended Deviation An unintended deviation is any error in the planned patient simulation, setup, treatment, or data entry in these processes. Any deviation from the planned course of treatment Any error in calculation Any missing or incomplete information Any failure to perform or follow required quality assurance and radiation safety policies or procedures Unintended deviations can be classified as: Pre or post-tx error A minor unintended deviation (Level 3-5) A significant unintended deviation (Level 1-2) A Recordable Event A Misadministration

Slide32: 

. Minor Unintended Deviations: 3rd Qtr. 1996 39% 9% 8% 7% 6% 5% 5% 4% 4% 4% 4% 4% 1% 0% 0% Data Entry: Daily Tx Record Process: Simulation Films Process: Patient Simulation Data Entry: ACCESS - Tx Field Def Tx of Patient: Port Films Data Entry: Tx Chart - Rx Data Entry: Tx Field Info Process: Block Fabrication Tx of Patient: Patient Beam Modifiers Process: Dose Calculation Data Entry: Patient Setup Doc QA: Missing or Late Radiation Safety: Missing or Late Tx of Patient: Patient ID Tx of Patient: Patient Setup

Slide33: 

TSUD - 2nd Qtr '96 TSUD - 3rd Qtr '96 0 20 40 60 80 100 120 140 160 180 Significant Unintended Deviations: 2nd & 3rd Qtr. 1996

Reported Misadministration Rate In Radiation Oncology: 

Reported Misadministration Rate In Radiation Oncology Published rates19 for reported misadministrations in therapeutic radiation oncology is 0.004 percent (4/100,000 administrations) based upon 20 treatments/patient for NRC regulated states only. Based upon internal NRC documents, it is speculated that the rate may be as high as 0.04 percent. 19 NRC memorandum dated March 8, 1993: Data based on information obtained from the American College of Radiology (Manpower Committee, Patterns of Care Study, and Commission of Human Resources). Additional reference from Institute of Medicine (Radiation in Medicine - A Need For Regulatory Reform), 1996.

Calculated Error Rates Paper-Based Model: 

Calculated Error Rates Paper-Based Model Based upon the total number of treatment fields delivered as recorded by R&V at 17 radiation oncology centers and the total number of unintended deviations self-reported by the system, excluding the initial two quarters for the “learning curve effect”, the overall average error rate for both minor and significant unintended deviations within the system was approximately 0.052% (5.2 in 10,000 patient treatments). The minor unintended deviation reporting rate for the same period was approximately 0.034%.

Measured vs Published Misadministration Rate Radiation Oncology: 

Measured vs Published Misadministration Rate Radiation Oncology The significant unintended deviation reporting rate that could lead to a misadministration was calculated to be approximately 0.018% (1.8 in 10,000 patient treatments).20 Based upon the model’s experience of one reported misadministration (having no deterministic or measurable effect) over 2 years, the measured misadministration rate was 0.017%. 20 Reporting rate is based on the number of significant interactions occurring in the treatment delivery process that could lead to a misadministration (criteria based on 10 CFR Part 35) vs the total number of treatment fields administered for 17 centers.

Measured vs Published Misadministration Rate Radiation Oncology: 

Measured vs Published Misadministration Rate Radiation Oncology When compared to what the NRC speculates is the actual misadministration rate of 0.04 (4 in 10,000), this rate is a factor of 2.35 lower. Though this program helped in minimizing the occurrence of misadministrations, the overall focus was to reduce the number and nature of all errors in the therapy process.

Cost Benefit Analysis: 

Cost Benefit Analysis After implementation of the QA/Medical Error Reduction Program, the 17 radiation oncology centers experienced a reduction of 326% in error rate from 3/96 to 12/97 (not including the “learning curve effect”): Direct cost savings of approximately $450,000 Direct & indirect cost savings of approximately $600,000

Cost Benefit Analysis: 

Cost Benefit Analysis Experience with the one reported misadministration that occurred at a center in Florida between 3/96 and 12/97 (with no measurable effect) resulted in a total direct cost (man-hours, travel, etc.) of approximately $25,000. Physician malpractice insurance premiums for the 17 oncology centers were reduced by 10%.

Summary of Results Paper-Based Model: 

Summary of Results Paper-Based Model Overall average error rate was 0.052% (SL 1 – 5) Calculated misadministration rate21 was 0.018% Actual misadministration rate was 0.017% NRC misadministration rate was 0.04% (a factor of 2.35 higher than actual misadministration rate) Reduced overall error rate by 326% over 21 months Direct cost savings of $450,000 Direct & indirect cost savings of $600,000 Other significant incidents averted by using program 21 Misadministration criteria based on definitions found in NRC 10CFR35.2, rev. 1996.

Other Center Studies Paper-Based Model: 

Other Center Studies Paper-Based Model Summary of Results - 1998 Oncology Company With 10 Freestanding Centers Three significant radiation treatment errors, that if left undetected would have required reporting to the State and notifying the referring physician and patient, were caught. A misadministration at one center, involving possible civil penalties and sanctions, was mitigated by the State by demonstrating that the error leading to the misadministration was isolated based on empirical data.

Other Center Studies Paper-Based Model: 

Other Center Studies Paper-Based Model Summary of Results - Calendar Year 2002 Cancer Center #1 Aside from the 1st quarter “learning curve”, total errors decreased by 70.5% (334 vs 99) between the 2nd and 3rd quarters. Total errors decreased by 27.3% (99 vs 72) between the 3rd and 4th quarters. The total decrease in errors between the 2nd and 4th quarters was 78.4% (334 vs 72). Cancer Center #2 Aside from the 1st quarter “learning curve”, total errors decreased by 66.4% (113 vs 38) between the 2nd and 3rd quarters. Total errors decreased by 18.4% (38 vs 31 between the 3rd and 4th quarters The total decrease in errors between the 2nd and 4th quarters was 72.6% (113 vs 31).

Lessons Learned With Paper-Based Model: 

Lessons Learned With Paper-Based Model Limitations Inefficient Time intensive Intrusive Complex industrial engineering model Requires paper trail Weaknesses Learning error codification system Triggering required regulatory actions Faxing of errors Tracking UDs Management review Trending and analysis Report generation Timely action Credible root cause analysis

Design of Software-Based Model: 

Design of Software-Based Model What is needed? Automated tracking of errors Non-intrusive data gathering Preset standardized gathering Immediate analysis of errors Short and long-term corrective actions Tracking and trending of errors Automated regulatory report launching

Design of Software-Based Model MERP Program Features: 

Design of Software-Based Model MERP Program Features Monitored Areas Clinical QA Radiation Safety Identification and Tacking of Errors Preset standardized error codes Classification of pre and post-treatment errors Assignment of severity levels (I - V) Designation of clinical significance Designation of significant unintended deviation "Near Miss" categorization Sentinel events (internal and JCAHO reportable) Instant analysis of patterns and trends Identification and Tacking of Violations Preset standardized unintended deviation codes Assignment of severity levels (I - V) Recordable events Misadministrations (medical events) Regulatory violations Possible regulatory violations Instant analysis of patterns and trends

Design of Software-Based Model MERP Program Features: 

Design of Software-Based Model MERP Program Features Step-By-Step Root Cause Analysis Determination of credible root cause analysis Identification of causal factors Identification of opportunities for improvement Action Plan Road Map Risk-reduction strategy Short-term corrective action Long-term corrective action Assignment of responsible individuals Patient Dose Error Calculation Wizard Calculates % error in daily, weekly & total doses Patient Dose Error Calculation Wizard (cont.) Automatically triggers levels for report generation JCAHO root cause analysis and action plans State regulatory notifications Review and Approval Queue action plan(s) for review and approval Accept or reject routine corrective action(s)

Design of Software-Based Model MERP Program Features: 

Design of Software-Based Model MERP Program Features Reports and Chart Generation Generate reports showing characterization of errors and corrective actions Show charts stratifying error types and severity levels Select time intervals for charting of data Audit Compliance Tool Use MERP to inspect regulatory performance Complies with State radiation safety requirement for annual review Meets State QMP rule for annual review

Design of Software-Based Model MERP Program Features: 

Design of Software-Based Model MERP Program Features Customization Features Customize and create data collection areas for performance improvement priorities Categories Subcategories Attributes Designate who reviews/approvals routine errors and corrective actions Assign which errors violate State requirements Designate severity levels, clinically significant, and significant unintended deviations Standards/Requirements Referenced by Code JCAHO 2005 patient safety standards show basis for question ACR and ACRO standards demonstrate benchmark for measuring performance CRCPD (Agreement State) recommended regulations (as of 9/04) show legal text

Conclusion: 

Conclusion Based on the experience gained from the clinical application of the paper-based model at over 42 centers throughout the country (29 described in this presentation), a software-based medical error reduction program (MERP) was developed. MERP provides a non-intrusive and efficient means to address medical error reduction in a systematic manner. Through implementation of MERP, errors that affect patient safety and/or result in regulatory violations can be minimized and often prevented for recurring. The initial results from the clinical application of MERP appear very promising. MERP ver.2.3 will now be demonstrated.